PortfolioHuman IVD · HPLC
UKGH-200
Glycohemoglobin Analyzer
Automated HPLC analysis with original-tube puncture sampling, continuous loading, and simultaneous IFCC and NGSP reporting. Three selectable modes — Routine, Variant and Thalassemia-related — use the same reagent system, which keeps inventory simpler. Load whole-blood samples, choose the mode, then review quality control, alarms and chromatograms before reporting. For professional use.
- Original-tube puncture sampling and continuous loading
- IFCC and NGSP results reported together
- Routine, Variant and Thalassemia-related modes on one reagent system
- LIS connection
Specifications
Published figures for this configuration. Confirm the reagent menu and the regulatory file for your market before you order.
| Test item | Glycated hemoglobin (HbA1c) |
|---|---|
| Method | High-performance liquid chromatography (HPLC) |
| Linear range | 3%–18% HbA1c |
| Sample types | Venous whole blood or fingertip peripheral blood |
| Sample volume | 5 μL |
| Sample capacity | 15 positions plus CAL1/QC1, CAL2/QC2 and STAT |
| Wavelength | 415 nm ± 5 nm |
| Analysis time | No more than 3 minutes |
| Reporting | IFCC and NGSP simultaneously |
| Connectivity | LIS |
| Dimensions | 460 × 452 × 390 mm |
| Net weight | 28 kg |
Supplied evaluation data
Figures reported with the instrument. They describe the stated evaluation, not a promise for every laboratory or every market.
- Accuracy
- Relative deviation within ±2% for stated reference materials
- Repeatability
- CV ≤ 1.7% in the stated range; carryover ≤ 0.6%
- Stability
- Relative deviation within ±1.8% over eight hours
- Correlation
- Reported coefficient r > 0.9900 versus an established comparator